top of page

Technical specifications

Model 2

Physical Specification

Dimensions: 327 mm x 310 mm x 493 mm (L x W x H): 664 mm x 600 mm x 1520 mm (with trolley)

Weight: Approximately 12.0 kg, Approximately

33.0 kg (with trolley)

Screen

Display Size: 18.3 Color active matrix TFT touch

Display Resolution (H) x (V): 1080 x 1980 pixels

Brightness: Adjustable

Ventilation Specifications

Patient Type: Adult, Pediatric, Neonate Invasive

Ventilation Modes:

VCV (Volume Control Ventilation)

PCV (Pressure Control Ventilation)

VSIMV (Volume Synchronized Intermittent

Mandatory Ventilation)

PSIMV (Pressure Synchronized Intermittent

Mandatory Ventilation)

CPAP/PSV (Continuous Positive Airway Pressure/Pressure Support Ventilation)

PRVC (Pressure Regulated Volume Control)

V + SIMV (PRVC + SIMV)

BPAP (Bilevel Positive Airway Pressure)

APRV (Airway Pressure Release Ventilation) Apnea Ventilation

Non-invasive Ventilation Modes:

CPAP (Continuous Positive Airway Pressure)

PCV (Pressure Control Ventilator)

PPS (Proportional Pressure Support)

S/T (Spontaneous and Timed)

 

Controlled Parameters

O2%: 21-100% (increments of 1%)

VT (Tidal Volume): Adult: 100-2000 mL (increments of 10 mL) / Pediatric: 20-300 mL /Neonate: 2-300 mL (increments of 1 mL)

f (Ventilation frequency): 1-80 bpm / Neonate:

1-150 bpm (increments of 1 bpm) fSIMV (Ventilation frequency in SIMV mode): 1-80 bpm / Neonate: 1-150 bpm (increments of 1 bpm)

I:E range: 4:1-1:10 (increments of 0.5)

Tinsp (Inspiratory time): 0.20-10 s (increments of 0.05 s)

Tslope (Time of Pressure Rising): Thigh 0-2.00 s (increments of 0.05 s)

Tlow: 0.2-30 s (increments of 0.1 s)

Tpause: 0.2-30 s (increments of 0.1 s) ΔPinsp: 5%-60 cm H2O (increments of 1 cm H2O), Off

ΔPsupp: 0-60 cm H2O (increments of 1 cm H2O)

Phigh: 0-60 cm H2O (increments of 1 cm H2O)

Plow: 0-45 cm H2O (increments of 1 cm H2O)

PEEP: 1-45 cm H2O (increments of 1 cm H2O), Off

Flow trigger: 0.5-15 L/min (increments of 0.1 L/min)

Pressure trigger: -10 to -0.5 cm H2O (increments of 0.5 cm H2O)

Exp% (Expiration termination level): 10-85% (increments of 5%), Auto

CPAP: 4-25 cm H2O (increments of 1 cm H2O)

EPAP: 4-25 cm H2O (increments of 1 cm H2O)

IPAP: 4-20 cm H2O (increments of 1 cm H2O) Rise time: 1-5 (increments of 1)

Ramp time: 5-45 min (increments of 5 min), Off

Min P (VS minimum IPAP): 5-30 cm H2O (increments of 1 cm H2O)

Max P (VS maximum IPAP): 6-40 cm H2O (increments of 1 cm H2O)

Max P (PPV maximum pressure limit):

5-40 cm H2O (increments of 1 cm H2O)

Max V (PPV maximum volume limit):

200-3500 mL (increments of 5 ml)

Max E: 0-100 cm H2O/L (increments of 1 cm H2O/L)

Max R: 0-50 cm H2O/L (increments of 1 cm H2O/L)

PPV%: 0%-100% (increments of 1%)

Apnea Ventilation

Vtvapnea: Adult: 100-2000 mL (increments of 10 mL) / Pediatric: 20-300 mL / Neonate: 2-300 mL (increments of 1 mL)

ΔPapnea: 5-60 cm H2O (increments of 1 cm H2O)

Fapnea: 1-80 bpm (increments of 1 bpm) Apnea

Tinsp: 0.20-10 s (increments of 0.05 s)

Sigh

Sigh Switch: On, Off

Interval: 20 s-180 min (increments of 1 s from 20 to 59 s, increments of 1 min from 1 to 180 min)

Cycles Sigh: 1-20 (increments of 1)

Δint.PEEP: 1-45 cm H2O (increments of 1 cm H2O), Off

 

Synchronized Tube Resistance Compliance

Tube Type: ET Tube, Trach Tube, Disable STRC Tube

I.D.: Adult: 5.0-12.0 mm (increments of 0.5 mm) / Pediatric: 2.5-8.0 mm (increments of 0.5 mm)

Compensate: 0-100% (increments of 1%)

Expiration Compliance Switch: On, Off

 

Monitored parameters

Numeric:

​

Real time Graphics:

Pressure-time waveforms: Paw-Volume Loop

Flow-time waveforms: Flow-time Loop Volumetime waveforms: Paw-Flow Loop

 

Technical Specifications

Control Accuracy

O2%: ±(3 vol.% + 1% of setting)

TV: ±(10 mL + 10% of setting) (BTPS) Tinsp: ±0.1 s or ±10% of setting, whichever is greater

I: E: 2:1 to 1:4: ±10% of setting, other range: ±15% of setting

f: ±1 bpm

fSIMV: ±1 bpm

Tslope: ±(0.2 s + 20% of setting)

PEEP: ±(2.0 cm H2O + 5% of setting) ΔPinsp: ±(2.0 cm H2O + 5% of setting) ΔPsupp: ±(2.0 cm H2O + 5% of setting) Phigh: ±(2.0 cm H2O + 5% of setting) Plow: ±(2.0 cm H2O + 5% of setting) Thigh: ±0.2 s or ±10% of setting, whichever is greater

Tlow: ± 0.2 s or ±10% of setting, whichever is greater

Pressure Trigger: ±(1.0 cm H2O + 10% of setting)

Flow Trigger: ±(1.0 L/min + 10% of setting)

Δint.PEEP: ±(2.0 cm H2O + 5% of setting) Exp %: ±10%

CPAP: ±(2.0 cm H2O + 5% of setting) EPAP: ±(2.0 cm H2O + 5% of setting) IPAP: ±(2.0 cm H2O + 5% of setting)

Rise time: -

Ramp time: ±1 s

Min P (VS minimum IPAP): ±(2.0 cm H2O + 5% of setting)

Max P (VS maximum IPAP): ±(2.0 cm H2O + 5% of setting)

Max P (PPV maximum pressure limit): ±(2.0 cm H2O + 5% of setting)

Max V (PPV maximum volume limit): ±15% of setting

Max E: -

Max R: -

Fapnea: ±1 bpm

ΔPapnea: ±(2.0 cm H2O + 5% of setting) Tvapnea: ±(10 mL + 10% of setting) (BTPS) Apnea Tinsp: ±0.1 s or ± 10% of setting, whichever is greater

 

Monitoring Accuracy

Airway pressure (Ppeak, Pplat, Pmean, PEEP, PAP,

EPAP): ±(2 cm H2O + 4% of the actual reading)

Tidal Volume (Tvi, Tve, TVe/IBW,TVe spn):

0 ml -100 ml: ± (10 ml+3% of the actual reading) (BTPS) / 100 ml - 4000 ml: ±(3 ml + 10% of the actual reading) (BTPS) Minute

Volume (MV, MVspn, Mvleak): ±0.3 L/min or ±8% of the actual reading, whichever is greater (BTPS)

Frequency (ftotal, fmand, fspn): ±5% of reading or ±1 bpm, whichever is greater

Inspired Oxygen (FiO2): ±(2.5 vol.% + 2.5% of the actual reading)

Resistance: 0 to 50: ±10 cm H2O/L/s Other range:

50% of the actual reading Compliance: 25% of the actual reading or ±10 ml/cm H2O, whichever is greater

RSBI: 0 to 999 1/(min*L): ± (3 1/(min*L)

± 15% of the actual reading)

WOB: -

NIF: ± (2 cm H2O + 4% of the actual reading) P0.1: ± (2 cm H2O + 4% of the actual reading) PEEPi: -

Rcexp: -

 

Alarm settings

Tidal Volume: High / Low

Minute Volume: High / Low

Airway pressure: High / Low

Frequency: High / Low

Inspired Oxygen (FiO2): High / Low

etCO2: High / Low

Apnea alarm time: 5-60 s

Trend

Type: Tabular, Graphic

Length: 72 hours

Content: Monitor Parameters,

Setting Parameters (Setting Ventilation mode and Parameters)

O2 Therapy

Controlled Parameters

O2%: 21-100% (increments of 1%)

Flow: 4-60 L/min

Controlled Accuracy

O2%: ±(3 vol.% +1% of setting)

Flow: ±(2 L/min +10% of setting) (BTPS)

 

Environmental specifications

Temperature: 5-40°C (operating); -20 to 60 °C (storage and transport, O2 sensor: -20 to 50 °C)

Relative Humidity: 10-95% (operating); 10-95% (storage and transport)

Barometric Pressure: 62-106 kPa (operating);

50-106 kPa (storage and transport)

Power Battery Backup

External AC power supply

Input voltage: 100-240 V

Input frequency: 50/60 Hz

Input current: 2.5 A Max

Fuse: T2.5 AH/250 V

Internal battery

Number of batteries: One or Two (Optional)

Battery type: Build-in Lithium-ion battery, 11.25 VDC, 6400 mAh

Battery run time: 3 hours (Powered by one new fully-charged battery in standard working condition) / 6 hours (Powered by two new fully-charged

battery in standard working condition)

 

Technical Specifications

Others

Communication interface: RS-232, Ethernet, USB port, CO2 analyzer connector

Gas supply: O2

(HPO) Oxygen connector: NIST (DISS optional)

Gas supply pressure: 280-600 kPa